![Stryker MMP200 Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/stryker/mmp200/mmp200_installation-and-service-manual_1385968063.webp)
63
S
Brake Screws
Brake Screws
Brake Screw
Brake Screw
Horizontal Arm Brake Screws
(EHA Only)
Warning
Disregarding the information concerning the adjustment of the brakes
could result in injury to the patient or physician.
Note
If the brakes require excessive tightening or frequent adjustment, refer to the sec-
tion called, “Adjusting the down tube Screws.”
1. Fill the hollow nose of the brass brake pads with Castrol Optipit grease (PN 0682-300-260) and
lubricate the faces of the brake pads with a thin layer of the grease.
2. Thread the brake screws into position using a 4mm hex wrench. Tighten the brake screws until
resistance is initially felt; this indicates the brake screws are engaged.
3. Tighten the brake screws:
•
Shoulder Screws:
1 revolution (360°) past the point of resistance
•
Elbow Screws:
¾ of a turn (270°) past the point of resistance
4. Rotate the extension arm to multiple positions throughout its range of motion to check for drift-
ing. If drifting is observed, increase the brake force by turning the brake screws clockwise until
the suspension no longer drifts.
5. Finally, tighten the brake screw an additional:
•
Shoulder Screws: 1 revolution (360°) to allow for brake wear
•
Elbow Screws: ¼ of a turn (90°) to allow for brake wear
Содержание MMP200
Страница 2: ......
Страница 4: ......
Страница 68: ...68 S 5 Ensure that the cover 5 and the retaining screws 4 are fixed securely Adjusting the Mounting Flange ...
Страница 106: ...106 S Front View of Connector ...
Страница 125: ... 125 S Schematics of Bladder Installation at Bearing 2 Raceway ...
Страница 140: ...140 S 18 Servicing the LED Refer to Visum LED Surgical Lights Service Manual 1004400195 for servicing instructions ...
Страница 148: ...148 S ...
Страница 149: ...Stryker Communications 1410 Lakeside Pkwy Flower Mound TX 75028 t 972 410 7100 www stryker com ...