10.5 Adjusting the Brakes
Once the center of gravity has been properly adjusted, you may need to tighten or loosen the tension
on the brakes.
Brakes
Brake location
To adjust the brake tension, insert a 3mm Allen
wrench into the socket head screw located on
the underside of the respective height adjust arm
and tighten the screw in a clockwise direction
or loosen in a counterclockwise direction. Only
tighten or loosen the screw slightly to determine
if it is sufficient to hold the monitor in place.
Adjust brake screws identically for both sides
at the same time (i.e., turn the right side one
quarter turn, then turn the left side one quarter
turn). Continue to adjust the brakes a little bit at
a time until they easily hold the monitor in place
throughout it’s range of motion.