110
S
c) One LED is the command to reset the keypad. If the reset is successful, the keypad will cycle.
Note: This command does not alter software programmed addressing.
d) Three LEDs is the command for Can-bus ID addressing. This sets the can-bus ID of the light. The
procedure is in step 5 below.
e) Six LEDs is the command for module testing. This is not used in the field.
5. E LED Can-bus ID Addressing (Self Addressing)
Can-bus ID addressing is software controlled and is done from the keypad service menu. All lights are
addressed for the installation configuration when they leave the factory (same as other E Series). Only
re-address when moving light heads or replacing a damaged light head or component. The addressing
DIP switches on the LED Module and Distribution Board are irrelevant once the self-addressing has been
completed. The change is stored in the software permanently.
a) Using the Addressing DIP switches on the keypad, set the ID for the corresponding light head to be
addressed.
Dip Switch Addressing:
Off-Off = Light 1, On-Off = Light 2, - Off-On = Light 3
b) Isolate all lights except the one being addressed from the can-bus circuit.
Note: There are several ways to isolate the lights from the can-bus circuit. They can be
disconnected at the suspension, at the tube, at a distribution board, etc. In most cases,
the suspension connections will be the easiest to access.
Содержание Berchtold CHROMOPHARE
Страница 1: ...CHROMOPHARE Surgical Lights Single and Combination Lights Service and Parts Manual ...
Страница 2: ......
Страница 21: ...16 S 5 7 ChromoView Monitor Supports Yoke 5 7 1 Single Monitor Support Arm 5 7 2 Dual Monitor Support Arm ...
Страница 64: ... 59 S ...
Страница 65: ...60 S ...
Страница 85: ...80 S 38 Plug in the programmed light control set 39 Tuck the wiring for the control pad into the cardanic ...