53
S
7.17 Horizontal Arm Maintenance
WARNING:
Before performing the following steps, the electricity for the Lights and/or
Monitors has to be powered OFF.
7.17.1
Replacing Height Compensation Tube (HCT) on Horizontal Arm
The HCT is a vertical down tube, available in varying lengths, used to put the spring arms are in an equal
horizontal plane.
Note: When replacing height compensation tube with Flat Panel arm, the Cables shall be
back pulled from the doc station.
1. Remove Light Head or Monitor Yoke per section 7.9.
2. Remove the Spring Arm per section 7.5.
3. Remove plastic cover (1) with round-headed screw (2).
4. Remove the special washer (2) with the round-headed screw (3) from the upper screw thread.
5. Remove the round-headed screw (1) from the upper screw thread.
6. Remove the height Compensation tube from the horizontal arm.
Содержание Berchtold CHROMOPHARE
Страница 1: ...CHROMOPHARE Surgical Lights Single and Combination Lights Service and Parts Manual ...
Страница 2: ......
Страница 21: ...16 S 5 7 ChromoView Monitor Supports Yoke 5 7 1 Single Monitor Support Arm 5 7 2 Dual Monitor Support Arm ...
Страница 64: ... 59 S ...
Страница 65: ...60 S ...
Страница 85: ...80 S 38 Plug in the programmed light control set 39 Tuck the wiring for the control pad into the cardanic ...