92
S
6. Tighten all screws on Ring A and install focus handle.
7. Clean face glass prior to installing light.
8. Reinstall the light head onto the spring arm per section 7.9.
9. Restore power to the system. Insure all modules illuminate and that all controls operate properly.
9.4 E550/650 (Halogen) and E655/805 (HID) Light Head Maintenance
HID Technology – E805/E655
HID Primary bulb
BRITe Halogen Backup bulb
BRITe Technology – E650/E550
BRITe Halogen Primary bulb
BRITe Halogen Backup bulb
9.4.1
Dome Removal and Replacement Exx0/Exx5
1. Secure power at circuit breaker prior to proceeding.
2. Using gasket removal tool gently insert flat side of tool under gasket and pry the dome up far enough to
get the tool under the dome. Gently rotate the light frame while moving the tool until the dome is able
to be extracted from the gasket.
3. If EndoLite option is installed (see section 9.4.2), disconnect the EndoLite socket wiring from the CPU
board connector J8.
4. Remove the dome assembly. If gasket is to be replaced, remove the old gasket now.