31
S
7.9.2
Flat Cardanic
WARNING:
Failure to install the locking segment WILL result in the Light Head or
Monitor Yoke falling.
Failure to secure the lock bushing with the screw may result in the locking segment
to disengage resulting in the Light Head or Monitor Yoke falling.
1. Remove the brake screw (1).
2. Turn the locking bushing (4) 180° exposing the shoulder screw (3).
3. Remove the shoulder screw (3).
4. Remove the Light Head.
5. With the replacement Light Head, align the contact pins inside the journal of the light bow.
Содержание Berchtold CHROMOPHARE
Страница 1: ...CHROMOPHARE Surgical Lights Single and Combination Lights Service and Parts Manual ...
Страница 2: ......
Страница 21: ...16 S 5 7 ChromoView Monitor Supports Yoke 5 7 1 Single Monitor Support Arm 5 7 2 Dual Monitor Support Arm ...
Страница 64: ... 59 S ...
Страница 65: ...60 S ...
Страница 85: ...80 S 38 Plug in the programmed light control set 39 Tuck the wiring for the control pad into the cardanic ...