96
S
E 550/650 Arrangement
E655/805 Arrangement
7. For E655/805, mark and remove the white wire from the top of the heat plate. The 2 red wires must be
disconnected at terminals 2 and 3 on the EVG board.
EVG E655
EVG E805
8. Remove heat plate screws and carefully remove heat plate to access the filter glass assembly.
9. Pull filter glass assembly up about 1 inch to expose the 6 pin wiring harness for the shutter motor.
10. Disconnect this connector and remove the filter glass from the reflector assembly. Match mark the
reflector assembly to indicate the position of the shutter motor assembly.
11. Inspect new filter glass damage and cleanliness ensure all shutters are straight and free for operation,
this can best be done by disconnecting the shutter motor resistor by removing the 2 screws holding it
to the assembly and manually exercising the shutters. Once testing is complete reinstall shutter motor
assembly onto filter glass assembly.