39
S
Note: Turning the height screw in the clockwise direction reduces the height and in a
counter clockwise direction increases the height.
7.11.3
Height Adjustment for NFC Spring Arms
1. Slide the plastic sleeve forward; exposing the adjustment slot.
2. Bring the spring arm to a position whereby the adjustment nut (2) is visible in the slot.
3. With a steel pin or equivalent (3) rotate the adjustment nut until the desired height limitation is
achieved.
7.11.4
Counter Balancing the Spring Arm for Lights and Flat Panels
1. Remove the cover cap(1) using a small screw driver. Adjustment screw for counterbalancing is beneath
the cover cap.
2. Put the spring arm into the closest position to be parallel with the floor.
3. Adjust the spring arm tension using the adjustment screw (2).
Note: Turning the height screw in the clockwise direction reduces the height and in a
counter clockwise direction increases the height.