63
S
5. Remove the three screws from the Endo light or jar cover and set them aside.
WARNING:
To prevent electrocution and burns, make sure the power to the light
system is turned off.
6. If you have an Endo light:
a. Unplug the 4 pin molex connector.
b. Remove the three screws from the internal plate and pull the plate away.
c. Disconnect the cable.
7. If the suspension has more than one arm, use an equipment lift to support the other extension arms.
WARNING:
To prevent injury or damage, do not proceed unless multi-arms are
supported by an equipment lift.
8. Use the castlenut driver to remove the jammer nut.