88
S
Module wiring locations on distribution board
4. Remove the 5 screws (4 corners and 1 in the center) holding the distribution board in place and lift the
distribution board out of the light head.
5. Place the new board with the same orientation and secure with the 5 screws removed in step 4 above.
Note: Grounding screw.