94
S
Parts List
1
64696 – Endo Bracke
2
64698 – Endo Cover Ring
3
66437 – Pivot for Endo Dome (Rivnut)
4
232 – M4 x 0.7 x 8 screw
5
64771 – Endo Socket
6
59833 – Endo Bulb 12V/20W
7
33 - Washer
8
39409 – M2.5 x 45 screw
9
62408 – Silicone Gasket
10
EndoLite label
1. The replace the bulb:
a) Remove the two fasteners (4), the cover ring (2) and the gasket (9).
b) Pull gently up on the bulb (6) to dislodge.
c) Install new bulb into socket (5).
d) Install the two fasteners (4), the cover ring (2) and the gasket (9).
2. To replace the socket assembly:
a) Remove the dome per section 9.4.1.
b) Remove the two screws (8) and washers (7) from the Endo bracket (1). Remove the socket
assembly (5).
c) Attach the new socket assembly with the same washers and screws.
d) Reinstall the dome per section 9.4.1.
9.4.3
Lower Glass Assembly
The lower glass assembly and the associated gasket are not field replaceable parts. Contact Technical
Support for a replacement light head.