Note:
Please refer to the disinfectant manufacturer's instructions-for-use for more detailed information regarding the
use of disinfectant solution, including proper rinsing techniques.
5. Dry the scope with a soft, lint-free, sterile cloth or clean, dry, oil-free compressed air (<5 psi). To dry out the
channel, flush it with clean, dry, oil-free compressed air (<5 psi).
Automated High-Level Disinfection
KARL STORZ and Medivators have completed a validation of the flow properties of the Medivators Reprocessing
Systems with the Flexible Intubation Video Endoscope. The scope is compatible with the following Medivators units
provided the correct hookups are used:
Reprocessor Model DSD/SSD and DSD Edge are used with hookup #DSD-110-HU0101 along with cleaning adaptor
11301CD1.
Reprocessor Model CER-1 (MV-1) and CER-2 (MV-2) are used with hookup # CLM-110-HU0080 and CLM-110-
HU0137 along with cleaning adaptor 11301CD1.
Reprocessor Model Advantage and Advantage Plus use hookup # 2-8-245HAN along with cleaning adaptor
11301CD1.
Recommended high-level disinfection agents for Medivators Reprocessing Systems:
Use CIDEX, CIDEX OPA, or Rapicide, or Rapicide OPA/28 for reprocessor models CER-1 (MV-1), CER-2 (MV-2), and
DSD/SSD.
Use Rapicide PA for reprocessor models DSD Edge and Advantage Plus.
Use only Rapicide for reprocessor model Advantage.
Reference for Cleaning and Sterilization
American National Standards Institute/Association for the Advancement of Medical Instrumentation: Flexible and
Semi-Rigid Endoscope Processing in Health Care Facilities: ANSI/AAMI ST91:2015.
Cleaning and Sterilization/Disinfection Instructions
43