Table 200
Accessories and cables which have been shown to comply with EN/IEC 60601-1-2:
Cable type
Shielded
Length (m)
Ferrite
Used for
Charger
No
2.0
No
Charging the internal
ET27-30-0004515
battery of C-MAC® monitor
8403 ZX
Table 201 /Table 1
Guidance and manufacturer’s declaration–electromagnetic emissions
The Flexible Intubation Video Endoscope 11302 BDX in conjunction with monitor 8403 ZX is intended for use in the
electromagnetic environment specified below. The customer or user of the system should ensure that it is used in
such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
Complies
The Flexible Intubation Video Endoscope 11302 BDX in
conjunction with monitor 8403 ZX uses RF energy only for its
CISPR 11
Class 1
internal function. Therefore, its RF emissions are very low and
are not likely to cause interference in nearby electronic
equipment.
RF emissions
Complies
The Flexible Intubation Video Endoscope 11302 BDX in
CISPR 11
Class B
conjunction with monitor 8403 ZX is suitable for use in all
establishments other than domestic and those directly
Harmonic emissions
Class A. Not
connected to the public low voltage power supply network that
IEC 61000-3-2
applicable
supplies buildings used for domestic purposes
Voltage fluctuations/flicker
Complies
emissions IEC 61000- 3-3
Electromagnetic Compatibility (EMC) Information
53