Inspecting the Surfaces
Warning:
The surface of the flexible sheath must be inspected visually and by carefully feeling it, to detect any
cracks, cuts, dents, foreign bodies, or other forms of damage.
CAUTION:
Do not squeeze the distal tip of the Flexible Intubation Video Endoscope too tightly.
Every time the Flexible Intubation Video Endoscope is used, you must first check that its surface, and above all the
surface of the shaft and distal tip, is in perfect condition and that there are no sharp edges. Intubation videoscopes
with sharp edges must not be used.
Inspect the connecting cord and connector for damage (such as breaks, cracks, twisted or squashed sections or
loose connections).
Checking the Deflection
Warning:
Actuate the deflection lever gently and carefully as far as the limit stop. The deflection lever must never be
operated with force or abruptly, to avoid injuring the patient or damaging the device! Move the deflection lever
smoothly and slowly!
Warning:
Before use, always check that the deflection lever and the distal tip connected to it function correctly.
Defective Flexible Intubation Video Endoscopes must not be used.
The distal end will deflect in accordance with the direction of actuation (forward or backward).
Installation and Operating
Instructions
24