6. Plug the instrument port with the Luer plug (P/N 29100-BK).
7. Draw a minimum of 50 mL of water into the syringe and flush back through the channel.
8. Remove the Luer plug (P/N 29100-BK) and attach a 60 cc syringe to the instrument port.
9. Draw a minimum of 50 mL of water into the syringe and flush back through the channel.
10. Remove the syringe from the instrument port and re-plug the port with the Luer plug (P/N 29100-BK).
11. Draw air into the 60 cc syringe and attach to rinse tube assembly (P/N 11301CDT). Holding the distal end of the
Flexible Videoscope out of the water, flush the 60 cc of air to expel all the water from the channel.
12. Discard the water and repeat with fresh water for a total of two (2) immersion/flush/rinses and air flushes.
Drying
1. Remove the Flexible Videoscope from the water.
2. Remove the cleaning adaptor (P/N 11301CD1) and cleaning cap (P/N 8401 YZ).
3. Discard all disposable cleaning accessories.
4. Dry the exterior with a fresh soft, lint-free cloth.
Inspection
1. Visually inspect the Flexible Videoscope for cleanliness. Do not proceed with sterilization or high-level disinfection if
residual soil is visible. Repeat all steps of the manual cleaning if residual soil is observed.
2. Prior to sterilization/high-level disinfection, inspect the Flexible Videoscope for signs of damage (cuts, dents, or
scratches) that may trap residual debris. If damage is found, the Flexible Videoscope should be immediately
tagged and labeled as damaged to prevent any further patient use. Refer to the “Endoscope Return Instructions”
for further instructions.
3. If no damage is found on visual inspection, clean the distal lens using a 70% isopropyl alcohol wipe. Wipe in a
circular movement.
4. Proceed to sterilization or high-level disinfection.
Cleaning and Sterilization/Disinfection Instructions
37