STERRAD Systems Sterilization
CAUTION:
STERRAD Sterilization may cause cosmetic changes to the Flexible Intubation Video Endoscope that do
not necessarily impact the functionality of the device.
CAUTION:
Place the pressure compensation cap (P/N Z11457) on the vent port before STERRAD sterilization.
CAUTION:
The Flexible Intubation Video Endoscope must be fully DRIED before loading into the STERRAD
sterilizer. Loads containing moisture may cause a cycle cancellation.
CAUTION:
Only use FDA-cleared polypropylene sterilization wraps. Do not use paper pouches or sterilization wraps
containing wood pulp or cotton.
1. Place the Flexible Intubation Video Endoscope in a STERRAD compatible instrument tray.
2. Place a STERRAD indicator strip in the tray.
3. Double wrap the sterilization tray with two single layers of FDA-cleared polypropylene wrap, using a sequential
double wrapping technique.
4. Load the STERRAD sterilizer, arranging the tray such that the vapor hydrogen peroxide can surround it. Do not
allow the tray to touch the wall of the sterilizer.
5. Start the STERRAD sterilizer using the 'FLEX' cycle for the 100NX system or the “Advanced” cycle for the NX
system.
6. For outside U.S. only: Use the "Long" cycle with the appropriate STERRAD Boosters for the 100S System. Place
adaptor 11301CD1 onto the flexible videoscope suction port to attach the STERRAD Booster Adaptor.
Note:
Please consult the STERRAD Sterilization System User Guide for detailed instructions for use of the STERRAD
Sterilizer, or contact Advanced Sterilization Products (ASP) customer service at (888) 783-7723.
Cleaning and Sterilization/Disinfection Instructions
40