Bedside Pre-cleaning (at point of use)
Water Quality Recommendation:
Utility water
1. Wipe the entire exterior of the flexible endoscope with a soft, lint-free, cloth, moistened with water, or a diluted
mild/neutral pH enzymatic cleaning solution (e.g. Enzol). This step is to remove gross debris from the exterior of the
flexible endoscope.
2. Ensure the instrument port is plugged during this process (e.g. use the Luer plug (P/N 29100-BK) or a gloved
finger to cover the port). Immerse the distal tip of the flexible endoscope in a basin of water. Depress the suction
valve (P/N 11301CE1) and suction (using a syringe or suction pump attached to the suction outlet) the channel with
a minimum of 100 mL (3.4 oz.) of water, or a diluted mild/neutral pH enzymatic cleaning solution (e.g. Enzol).
3. Disconnect the flexible endoscope from the monitor.
4. Inspect the vent port and the video pin connector area removing any visible debris on the external surface with a
70% isopropyl alcohol wipe.
5. Carefully place the flexible endoscope into a covered tray or container and transport the scope to the
decontamination area. Ensure the flexible endoscope is kept protected during transport.
6. Dispose of all valves and plugs.
Note:
Complete cleaning of the patient-used flexible endoscope should be started
within two hours
of the bedside
pre-cleaning. If transit time is greater than
2 hours
, ensure that additional manual cleaning (see Manual Cleaning
Section) is performed if visible residual debris is still present.
Cleaning and Sterilization/Disinfection Instructions
31