Sterilization Methods
Ethylene Oxide (EtO) Sterilization
1. Remove the cleaning cap 8401 YZ on the video connecting cord before sterilization.
2. Place the red pressure compensation cap (P/N Z11457) on the vent port before EtO gas sterilization.
3. Place the flexible endoscope in an FDA-cleared EtO gas sterilization tray.
4. Wrap the sterilization tray with two single layers of FDA-cleared polypropylene EtO gas-compatible sterilization
wrap, using a sequential double wrapping technique.
5. Process in 100% Ethylene Oxide Sterilizer under the following parameters:
CONDITIONING PARAMETERS
Temperature:
55° C (131° F)
Humidity:
70% RH
Conditioning Dwell Time:
30 minutes
STERILIZATION PARAMETERS
Temperature:
55° C (131° F)
Humidity:
70% RH
Humidity Dwell Time:
30 – 45 minutes
EtO Gas Concentration:
725 ± 30 mg/L
EtO Gas Exposure Time:
60 minutes
AERATION PARAMETERS
Time:
12 hours
Temperature:
51 – 59° C (120° –130° F)
Cleaning and Sterilization/Disinfection Instructions
39