Manual High-Level Disinfection
Warning:
High-level disinfection (minimum requirement) is recommended ONLY for "semi-critical" instruments which
only come into contact with mucous membranes or minor skin breaches.
CAUTION:
Remove the pressure compensation cap (P/N Z11457) from the vent port before immersing in liquids.
CAUTION:
Place the 8401 YZ protective cap on the video connecting cord prior to immersion in any liquids.
The Flexible Intubation Video Endoscope may be chemically disinfected using CIDEX, a high-level disinfectant
containing 2.4% concentration of glutaraldehyde, CIDEX OPA, a 0.55% concentration of
ortho
-phthalaldehyde, or
Resert XL HLD, a 2.0% concentration of hydrogen peroxide.
1. Prepare the disinfecting solution for use:
a. CIDEX: Activate the glutaraldehyde solution by adding the entire contents of activator vial to the solution in the
container. Shake well. Activated solution immediately changes color to green, thereby indicating the solution is
ready to use. Use CIDEX Solution Test Strips to verify the solution is above the minimum recommended
concentration (MRC). Test the solution prior to each use. Do not use activated solution beyond stated 14 day
reuse life. Record the date of activation and expiration on the container.
b. CIDEX OPA: No activation is necessary. Use CIDEX OPA Solution Test Strips to verify the solution is above the
minimum recommended concentration (MRC). Test the solution prior to each use. Record the date the solution
was poured out of the original container.
c. Resert XL HLD: No activation is necessary. Use Verify Chemical Monitoring Strip for Resert XL HLD Solution to
confirm hydrogen peroxide concentration before each use. Record the date the solution was poured out of the
container.
2. Place the Flexible Intubation Video Endoscope into a plastic container containing the disinfection solution. Ensure
that the scope is fully immersed and remove any air bubbles adhered onto the surfaces. Fill all working channels
with the disinfection solution using a syringe.
3. Use the following KARL STORZ validated conditions to achieve high-level disinfection:
a. CIDEX: Immerse for 45 minutes at 25° C (77° F)
b. CIDEX OPA: Immerse for 12 minutes at 20° C (68° F)
c. Resert XL HLD: Immerse for 8 minutes at 20° C (68° F)
4. After disinfection is complete, remove the flexible intubation video endoscope from the disinfection solution and
completely immerse in a large volume of sterile water (e.g. 2 gallons). Flush the channels with water to remove
disinfection solution. Keep immersed for one minute. Discard the water and repeat with fresh sterile water for a total
of three immersion rinses.
Cleaning and Sterilization/Disinfection Instructions
42