3 Design and functional description
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
63
Parameter
Unit
Lower limit
Upper limit
PAW
(VC mode)
mbar
-
Pplat+10 measurement
value
maximum 35 mbar, no
less than PEEP+5
PAW
(PC mode)
mbar
-
Pinsp + 10
maximum 35 mbar
PAW
(PC mode+PRVC)
mbar
No Auto button
available
No Auto button
available
PEEP
mbar
Set value -2-
Set value + 2
VTe
mbar
−30%
+30%
EtCO
2
mbar
−10%
minimum
35 mmHg
+10%
maximum
45 mmHg
Tab. 8: Automatic alarm limits
If configured in the setup menu, the alarm limits for FiO
2
, PAW LOW
and PEEP are always set automatically.
Parameter
Unit
Lower limit
Upper limit
O
2
%
Set value −10
minimum 21%
Set value +10
maximum 100%
PEEP
mbar
-
Set value + 5