
5 Preparation for use
122
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
CAUTION
The measurement of the minute volume (MV) may be affected by the medication
nebulization.
CAUTION
Remove the medication nebulizer after use. Unintended nebulization may put the
patient at risk and impair breathing.
CAUTION
The
EVE
TR
features built-in automatic compensation. As a result, the Aerosol button
(see chapter 3.1.3) should be pressed only when a medication nebulizer is connected.
Otherwise, the resulting breathing volume is too low.
NOTE
Aerosol nebulization is only possible if O
2
is connected to the ventilator.
NOTE
Aerosol nebulization is automatically synchronized with inspiration.
NOTE
In volume-controlled mode, the volume is automatically compensated for during
nebulization.
For medication nebulization, install the nebulizer as shown in Fig. 71 and
fill according to the operating manual provided with the nebulizer.
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...