
11 Care and maintenance
224
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
11.7.2
Test lung Neo with tube adapter
Instructions
Place of use
Remove surface dirt with a disposable cloth/paper towel
Storage and transport
No particular requirements
Cleaning preparations
1.
Remove the connection tube (3) from the test lung (1) together with
the tube adapter (4) and tube sleeve (2).
2.
Remove the tube adapter and tube sleeve from the silicone tube.
Fig. 114: Test lung with tube adapter
Cleaning: automatic
1.
Connect the silicone tube to the injector rail of the cleaning and
disinfection equipment.
2.
Position the test lung to ensure that the cavity will be fully rinsed and
that no spots are missed. Excess water must be able to run off freely.
3.
Place the tube sleeve in a small parts container.
4.
Load the automated disinfection equipment according to the
manufacturer’s instructions for use and start the program for
anesthesia materials.
5.
Dry the disinfected parts (in a drying cabinet if necessary), unless
this happens inside the automated disinfection equipment.
Cleaning: manual
Disinfection
Drying
1.
Immerse the individual parts in ready-to-use disinfectant solution.
All of the parts must be fully covered by the disinfectant.
2.
Move the parts around in the solution several times until all air
bubbles have been removed.
3.
After the dwell time, the disinfectant must be completely rinsed off
using cold water (at least drinking water quality).
4.
Dry the parts thoroughly.
5.
After completion, check the parts for visible external residual
contamination or disinfectant. If necessary, repeat the manual
cleaning and disinfection process.
Maintenance
1.
Check for residual water and carefully purge with sterile compressed
air if necessary.
2.
Connect the tube adapter and tube sleeve to the silicone tube.
3.
Connect the connection tube to the test lung together with the tube
adapter and tube sleeve
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...