GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Flow/
volume
Flow sensor
Dead space
Children
2.7 ml
Adults
11 ml
Preterm infants and newborns
(PNT B)
0.6 ml
Proximal measurement of flow (pressure difference) and
pressure.
CO
2
IRMA™ airway adapter
Dead space
Preterm infants and newborns ≤ 1 ml
Adults/children
≤ 6 ml
HME filter
Children
Air filter volume
8 ml
Connections
15f patient-side
15m machine-side
Luer lock
Pressure drop at
5 l/min
7.5 l/min
10 l/min
0.48 cmH
₂
O
0.63 cmH
2
O
0.78 cmH
₂
O
HME filter
Adults
Air filter volume
50 ml
Connections
22m/15f patient-side
22f/15m machine-side
Luer lock
Pressure drop at
50 l/min
0.2 kpa
O
2
(EVE
TR
)
Consumption-free measurement
Sensors