1 General information
24
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Helicopter bracket
Dimensions
426 x 360 x 150 mm (W x H x D)
Weight
3.3 kg
Connections:
12 – 28 V DC
Carry system without gas cylinder and pressure regulator
Dimensions
470 x 335 x 260 mm (W x H x D)
Weight
2.7 kg
Supply pressure
High pressure
O
2
280 – 600 kPa
Flow rate
Max. 200 l/min
Gas consumption
Conventional (see chapter 5.1.2 for more
information about O
2
gas consumption)
Aerosol
Approx. 7 l/min
at 2.8 bar
The gases must be dry and free of oil and dust.
PC-CMV
Pressure-controlled continuous mandatory ventilation
PC-SIMV
Pressure-controlled synchronized intermittent mandatory
ventilation
PC-ACV
Pressure-controlled assist/control ventilation
PC-ACV+
Pressure-controlled assist/control ventilation with
expiration trigger
CPAP
Spontaneous respiration with continuous positive airway
pressure
DUOPAP
Ventilation support at two different pressure levels
VC-CMV
Volume-controlled continuous mandatory ventilation
VC-SIMV
Volume-controlled synchronized intermittent
mechanical ventilation
nPC-CMV
Non-invasive pressure-controlled mandatory ventilation
nPC-SIMV
Non-invasive pressure-controlled synchronized
intermittent mandatory ventilation
Gas supply
Available
ventilation modes
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...