8 CO2 measurement (optional)
180
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Fig. 103: Connecting the socket of the airway adapter to the ETT tube
NOTE
The use of an HME filter is recommended! In this case, place the measuring probe
between the HME filter and the ventilator. The HME filter protects the airway adapter
from secretions and water vapor so that the adapter does not have to be replaced.
NOTE
The measured gas values should be checked regularly for accuracy by conducting
comparative measurements. The gas section should be tested annually.
NOTE
After inserting the probe, a leakage test of the
EVE
TR
patient system is recommended.
WARNING
The airway adapter must be replaced if moisture or condensation collect on the inside.