11 Care and maintenance
226
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
11.8
Treatment table
CAUTION
The processes for the treatment of the medical device described in this table are
recommendations only. Always observe the individual procedural instructions given
by the responsible hygiene officer.
WARNING
In case of infectious patients, all reusable parts of the patient system carrying
breathing gas must also be sterilized.
System components Recommended
treatment intervals
Automated
thermal
treatment
Manual
treatment
Sterilization
Case, power cord,
connection tubes
After every patient/
weekly
No
Surfaces
No
Touchscreen
After every patient/
weekly
No
Surfaces
No
Docking station
After every patient/
weekly
No
Surfaces
No
Ambulance bracket
If contaminated
No
Surfaces
No
Helicopter bracket
If contaminated
No
Surfaces
No
Carry system
If contaminated
No
Surfaces
No
Accessories bags
(carry system)
If contaminated
No
Clean manually
using a brush
No
Adult test lung
After every patient/
weekly
Observe the manufacturer’s operating
manual!
Test lung Neo with
tube adapter
After every patient/
weekly
Yes
Yes
Yes
Reusable
expiration valve
After every patient/
weekly
Yes
Yes
Yes
Reusable patient
tube system
After every patient/
weekly
Yes
Yes
Yes
Type B
pneumotachograph
After every patient/
weekly
Yes
Yes
Yes
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...