10 Troubleshooting
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
201
Battery 2 temp. low
MP
TA
Battery 2
undercooled
Check battery 2, replace
if necessary
Do not operate the
device at temperatures
outside ambient
conditions
(see chapter 1.9).
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Apnea
MP
PA
No inspiration
within the set apnea
alarm limits
Check the patient's
condition!
Does the patient actually
exhibit apnea?
Switch to mandatory
ventilation, if necessary
Check the flow sensor
Check the alarm limits
Respiratory rate high
(Alarm delay: 15 s)
MP
PA
Patient is breathing
at a respiratory rate
above the alarm
limit setting
Check the patient's
condition!
Check the alarm limits
Check the trigger
threshold and correct if
necessary
Airway pressure low
MP
PA
Volume
measurement
outside specified
accuracy
Flow sensor clogged
Water buildup in flow
sensor tube
Blower failure 1
MP
TA
Internal malfunction
of the O
2
valve
Check the measured and
set the value for O
2
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Blower failure 2
MP
TA
Malfunction of the
internal flow sensor
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Replace CO
2
adapter
MP
TA
CO
2
adapter
malfunction or
contamination
Replace the CO
2
adapter
CO
2
sampling line
clogged
MP
TA
Sampling line is
contaminated or
clogged
Check the sampling line
Replace the sampling
line
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...