10 Troubleshooting
200
GA-070T-0518V2.7-HAO-EN
Fritz Stephan GmbH
Battery low
MP
TA
Battery voltage low
Check the mains supply
and charge the battery
Have a spare battery at
the ready
If the battery cannot be
charged, contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Battery 1 temp. high
MP
TA
Battery 1 has
overheated
Check battery 1, replace
if necessary
Keep the device away
from direct sunlight or
other heat sources.
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Battery 2 temp. high
MP
TA
Battery 2 has
overheated
Check battery 2, replace
if necessary
Keep the device away
from direct sunlight or
other heat sources.
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Battery 1 temp. low
MP
TA
Battery 1
undercooled
Check battery 1, replace
if necessary
Do not operate the
device at temperatures
outside ambient
conditions
(see chapter 1.9).
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...