10 Troubleshooting
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
203
Leakage
(Alarm delay: 3 s)
HP
PA
If the inspiratory
tidal volume is 80%
more than the
expiratory tidal
volume after three
breaths
Check the patient's
condition
Check the tube system
Check the ETT
Fan failure
MP
TA
Fan malfunction
Check the fan, replace if
necessary
Low O
2
pressure
MP
TA
Oxygen delivery
pressure from the
gas cylinder or
central gas supply
too low (< 2.7 bar)
Check the oxygen
delivery pressure and
replace the O
2
cylinder if
necessary
Check the oxygen
pressure in CGS
O
2
high
(Alarm delay: 30s)
MP
PA
Calculated O
2
value
is above the
automatic alarm
limit value
Compare the set O
2
value
with the calculated value
In case of larger
deviations, contact F
RITZ
S
TEPHAN
G
MB
H
customer service
O
2
low
(Alarm delay: 30s)
MP
PA
Calculated O
2
value
is below the
automatic alarm
limit value
Compare the set O
2
value
with the calculated value
In case of larger
deviations, contact F
RITZ
S
TEPHAN
G
MB
H
customer service
Parameter error
MP
TA
Internal
communication
error
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...