3 Design and functional description
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
59
Alarms
3.2.3.3
The "Alarms" sub-menu shows the »Alarm History« and all measured
values that are monitored by alarm limits.
Pressing the »Alarm History« field or the alarm indicator
(see chapter 3.2.8) opens a list of the seven most recent alarms.
Fig. 18: "Alarm History" sub-menu
1 Time and date when the alarm
occurred
3 Alarm text
4 Active alarms
2 Alarm category
5 Inactive alarm
The system shows the date and time when the alarm occurred, the
category (P = patient alarm, T = technical alarm), and the alarm text.
The color indicates the priority (red = HP alarm; yellow = MP alarm;
white = alarm no longer active).
All active and inactive alarms are stored internally. However, the alarm
history only shows the 7 most recent alarms in chronological order. Once
the cause of the alarm has been eliminated, the message remains, but is
no longer highlighted in color. If an inactive alarm recurs, it becomes
current and appears in the alarm history.
Manually deleting an inactive alarm from the alarm history moves the
next internally stored alarm into the on-screen alarm history.
Alarms
Alarm history
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...