10 Troubleshooting
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
199
Powersupply?!
HW
Device has been
disconnected from
the external energy
supply
If possible, connect the
device to the external
energy supply
No device error if the
device was deliberately
disconnected from the
energy supply (undocked
from docking station,
device bracket for patient
transport). Notification
can be acknowledged
with the "Alarm
Suppression" button.
Nebulizer inactive/
check O2!
HW
Oxygen supply not
connected or
malfunctioning.
Check oxygen supply,
connect if necessary
Maintenance required
HW
Maintenance
interval reached
Perform service
Contact
F
RITZ
S
TEPHAN
G
MB
H
customer service
VTe high
HW/MP
PA
The inspiratory
breath volume is
above the upper
alarm limit for two
breaths
Escalates to MP
after 60 s
Check the patient's
condition
Check the settings and
adjust if necessary
Check the alarm limits
and adjust if necessary
VTe low
HW/MP
PA
The expiratory
breath volume is
below the lower
alarm limit for three
breaths
Escalates to MP
after 60 s
Check the patient's
condition
Check the settings and
adjust if necessary
Check the alarm limits
and adjust if necessary
Check the tube system
for leakage or
disconnected parts
Содержание EVETR
Страница 1: ...EVETR Transport Ventilator Instructions for use...
Страница 32: ......
Страница 40: ......
Страница 86: ......
Страница 106: ......
Страница 124: ......
Страница 192: ......
Страница 194: ...9 Functional description 194 GA 070T 0518V2 7 HAO EN Fritz Stephan GmbH Fig 108 EVETR pneumatics design...
Страница 210: ......
Страница 234: ......
Страница 240: ......
Страница 246: ......
Страница 253: ...17 Notes Fritz Stephan GmbH GA 070T 0518V2 7 HAO EN 253 17 Notes...