11 Care and maintenance
Fritz Stephan GmbH
GA-070T-0518V2.7-HAO-EN
223
Fig. 113: Connecting the pressure measurement lines on the reusable
flow measurement adapter
3
Connection 1
for proximal pressure measurement line
4
Connection 2
CAUTION
Ensure that all pressure measurement lines are mounted in the correct position. Any
confusion between different pressure measurement lines may result in inverted
measurements. If this occurs, the measurement results for tidal volumes and minute
volumes will be incorrect.
Before using the flow sensor, leak and functional checks must be
performed (see chapter 7 of the operating manual).
Function test