28
Specification Guide
SR620 Universal Time Interval Counter
PHASE MODE
In phase mode the measurement resolution and error are given by:
N
= number of samples averaged
note: the gate time is 10ms in internal mode
Resolution
=
±
0
.
001
°
±
360
(
25
ps
)
2
+
(
gate
time
×
short term stability )
2
+
2
×
(
trigger jitter
)
2
period
2
×
N
1
+
phase
×
period
360
×
gate time
2
Error
=
±
resolution
±
(
timebase
error
×
time
interval )
±
start
trigger level error
±
stop
trigger level
error
±
0
.
5ns
timebase error
×
period
±
1
×
10
−
8
×
period
×
360°
Graph 6 shows the SR620's single-shot phase resolution as a function of frequency. The resolution may be
increased by averaging.
Graph 6: Single-Shot phase resolution vs.
frequency.
COUNT MODE
The resolution and error for count mode are:
Resolution = ±1 count
Error = ±1 count
Содержание SR620
Страница 2: ...SR620 Universal Time Interval Counter...
Страница 6: ...iv Table of Contents SR620 Universal Time Interval Counter...
Страница 8: ...vi Safety and Preparation for Use SR620 Universal Time Interval Counter...
Страница 12: ...x Specifications SR620 Universal Time Interval Counter...
Страница 58: ...42 Programming Commands SR620 Universal Time Interval Counter...
Страница 72: ...56 Programming Examples SR620 Universal Time Interval Counter...
Страница 76: ...60 Troubleshooting Tips SR620 Universal Time Interval Counter...
Страница 82: ...66 Performance Test SR620 Universal Time Interval Counter...
Страница 90: ...74 Calibration Procedure SR620 Universal Time Interval Counter...
Страница 102: ...86 Circuit Description SR620 Universal Time Interval Counter...
Страница 124: ...108 Parts List SR620 Universal Time Interval Counter...