Specifications
vii
SR620 Universal Time Interval Counter
Specifications
Functions
Time Interval, Pulse Width, Rise and Fall Times, Frequency, Period, Phase,
and Event Counting.
Measurement statistics (mean, min, max, standard deviation or Allan
variance) and graphics are available in all modes of operation.
Time Interval
Time, Width, Rise and Fall Times
Range
-1000 to +1000 s in +/- TIME mode; -1 ns to +1000 s in all other modes
Trigger Rate
0 to 100 MHz
Least Significant Digit
4 ps single sample, 1 ps with averaging
Resolution
(((25 ps typ [50 ps max] )2+ (0.2 ppb x Interval)2) / N)1/2 rms
(((25 ps typ [50 ps max])2 + (0.05 ppb x Interval)2) / N)1/2 rms (Opt 01)
Error
< ±(500ps typ [1 ns max] + Timebase Error x Intrigger error)
< ±(50ps typ[100ps max] + Timebase Error x Interval) (REL)
Arming
Modes
+TIME
Stop
is
armed
by
Start
+TIME EXT
Ext arms Start
+TIME EXT HOFF Leading EXT edge arms Start, trailing EXT edge arms
Stop.
±TIME
Armed
by
Start/Stop
pair
±TIME CMPL
Armed by Stop/Start pair
±TIME EXT
Armed by EXT input edge
EXT arming may be internally delayed or scanned with respect to the EXT
input in variable steps. The step size may be set in a 1,2,5 sequence from 1
µs to 10 ms. The maximum delay is 50,000 steps.
Display
16 digit fixed point with 1 ps LSD
Sample Rate
For a sample size of N, the total sample time is N x(800 µs + measured time
interval) + Calculation time.
The calculation time occurs only after N measurements are completed and
varies from zero (N=1, no graphics, binary responses) to 5 ms (N=1, no
graphics) to 10 ms (display mean or std dev) to 60 ms (display histogram).
Frequency
Range
0.001 Hz to 300 MHz via comparator inputs.
40 MHz to 1.3 GHz via internal UHF prescalers.
RATIO A/B range: 10
-9
to 10
3
Error
< ± ((100ps typ [350 ps max])/Gate + Timebase Error ) x Frequency
Gates
External, 1 period, 1ms to 500s in 1,2,5 sequence. Gates may be externally
triggered with no delay. Gates may be delayed relative to an EXTernal
trigger. The delay from trigger may be set from 1 to 50,000 gate widths.
Display
16 digit fixed point with LSD = Freq x 4ps/Gate. 1 µHz max. resolution (1
nHz with x1000 for frequencies
≤
1 MHz)
Period
Range
0 to 1000 seconds. RATIO A/B range: 10
-9
to 10
3
Error
< ±((100 ps typ [350ps max])/Gate + Timebase Error) x Period
Gates
Same as frequency
Display
16 digit fixed point with LSD = 1 ps (1 fs with x1000 for periods
≤
1 s)
Phase
Phase = 360 x (Tb - Ta) / Period A
Range
-180 to +180 degrees (0 to 100 MHz frequency)
Resolution
(25ps x frequency x 360 + 0.001) degree
Содержание SR620
Страница 2: ...SR620 Universal Time Interval Counter...
Страница 6: ...iv Table of Contents SR620 Universal Time Interval Counter...
Страница 8: ...vi Safety and Preparation for Use SR620 Universal Time Interval Counter...
Страница 12: ...x Specifications SR620 Universal Time Interval Counter...
Страница 58: ...42 Programming Commands SR620 Universal Time Interval Counter...
Страница 72: ...56 Programming Examples SR620 Universal Time Interval Counter...
Страница 76: ...60 Troubleshooting Tips SR620 Universal Time Interval Counter...
Страница 82: ...66 Performance Test SR620 Universal Time Interval Counter...
Страница 90: ...74 Calibration Procedure SR620 Universal Time Interval Counter...
Страница 102: ...86 Circuit Description SR620 Universal Time Interval Counter...
Страница 124: ...108 Parts List SR620 Universal Time Interval Counter...