A:28
General instrument functions
A
Technical specifications
Instrument, general
Housing: ABS / PC / TPE
Protection class:
IP65
Dimensions:
306 x 108 x 71 mm
(12.4” x 4.3” x 2.8”)
Weight:
860 g (30.3 oz.)
Keypad:
sealed, snap action
Display:
TFT colour, 240 x 320 pixels,
3.5 inch, adjustable backlight
Main processor:
400 MHz ARM
®
Memory:
256 MB RAM, 512 MB Flash
SD memory card:
1GB
Operating system:
Microsoft Windows
®
CE
DSP processor:
375 MHz floating point
Dynamic range:
up to 120 dB,
24 bit A/D converter
Communication:
USB 2.0
Input channels:
1 x VIB, 1 x SPM,
1 x rpm , 1 x temperature
Input/output, headset: 3.5 mm stereo plug
Input/output,
communication:
Mini USB
Power supply:
rechargeable Lithium-Ion
battery pack 5200 mAh or
power adapter
Battery power:
for min. 18 hours normal use
(at 20°C)
Charge time:
< 4 hours
Operating temperature: -20 to 55 °C (-4 to 122 °F)
non condensing
Charging temperature: 0 to 45 °C (32 to 113 °F)
General features:
language selection, customized
view, voice recording, battery
status idication, transducer line
test, metric or imperial units
Meas. point
identification:
NFC transponder for
communication with CondID
TM
tags, read distance max. 50 mm
(2 inch)
SPM input
Shock pulse channels: 1
Measuring techniqes: SPM HD, dBm/dBc, LR/HR and
LR/HR HD
Transducer line test:
TLQ test
Transducer types:
SPM 40000, 42000, 44000,
probe transducer and quick con-
nector transducer
VIB input
Vibration channels:
1
Frequency range:
0 (DC) to 20 kHz
Resolution:
max. 12 800 lines
Transducer line test:
bias voltage check
Transducer types:
SPM SLD or IEPE* (ICP) type
with voltage output < 24 Vpp, or
DuoTech accelerometer. Trans-
ducer supply of 2.5 mA can be
set On/Off
User defined transd.:
e.g. current clamp, pressure and
flow sensors with voltage output
Measuring techniqes: ISO 2372, ISO 10816, FFT with
symptoms, EVAM vibration
analysis, HD Enveloping,
balancing
RPM input
Channels:
1, RPM or temperature
Measuring range:
1 to 150 000 pulses/min.
Resolution:
1 pulse
Accuracy:
± (1 pulse + 0.1% of reading)
Transducer type:
SPM TTP 10, TTL-pulses,
Key phasor
®
, proximity switch
NPN/PNP
Output:
TTL signal for controlling
stroboscope and 12 VDC supply
Temperature measurement
Temp. range (TTP10) : -20 to 300 °C (-4 to 572 °F)
Accuracy (TTP10) :
± 2.5 °C
Transducer type:
SPM TTP 10
Stethoscope
Settings:
Filter, volume and gain
Transducer types:
Shock pulse and vibration
transducers
Содержание leonova emerald
Страница 1: ...User guide PORTABLE EFFICIENCY ...
Страница 2: ......
Страница 6: ...A 2 General instrument functions A ...
Страница 36: ...B 2 General measurement functions B ...
Страница 74: ...C 2 Measurement techniques always included B C ...
Страница 88: ...D D 2 Shock pulse measurement ...
Страница 122: ...D D 36 Shock pulse measurement ...
Страница 124: ...E E 2 Vibration measurement ...
Страница 150: ...E E 28 Vibration measurement ...
Страница 152: ...F F 2 Rotor balancing ...
Страница 170: ...F F 20 Rotor balancing ...