IM-P198-05
MI Issue 2
32
R current limit - 10 k
DC power
DC common
Freq. Out +
Freq. Out -
+ Pwr Bklt
- Pwr Bklt
Freq. Out voltage = +V
Select resistor so that current through pulse
40 mA
3.16 Frequency output connections
The frequency output is used for a remote counter. It can be scaled to output a 1 to 10 kHz
signal proportional to mass or volume flow, temperature, pressure or density.
The frequency output requires a separate 5 to 36 Vdc power supply; however, there are
current and power specifications that must be observed.
The output can conduct a current up to 40 mA and can dissipate up to 200 mW. The output
is isolated from the meter electronics and power supply.
Fig. 19 Non-Isolated frequency output using external power supply
3.17 Optional backlight connection
The loop power meter has an optional backlight connection provided. It is intended to be
powered by a separate 12 to 36 Vdc at 35 mA max. power supply or by the pulse power input.
Both options are shown below.
Fig. 20 Backlight using external power supply
12 to 36 VDC
25 mA max.
+ Pwr Bklt
- Pwr Bklt
12 to 36 Vdc
25 mA max.
Fig. 18 Isolated pulse output using external power
+V
R current limit - 10 k
Freq. Out +
Freq. Out -
Pulse voltage = +V
Select resistor so that current
through pulse
40 mA
dc power
dc common
Freq. Out +
Freq. Out -
Freq. Out +
Freq. Out -
Freq. Out voltage = +V
Select resistor so that current through pulse
40 mA
R current limit - 10 k
Pulse voltage = +V
Select resistor so that current
through pulse
40 mA
+ Pwr Bklt
- Pwr Bklt
+ Pwr Bklt
- Pwr Bklt
Содержание RIM20
Страница 49: ...IM P198 05 MI Issue 2 49 Run mode screens Press Exit to return to Run mode Use keys to access each item...
Страница 50: ...IM P198 05 MI Issue 2 50 4 3 Using the set up menus Run mode screens...
Страница 52: ...IM P198 05 MI Issue 2 52 4 5 Output menu...
Страница 56: ...IM P198 05 MI Issue 2 56 4 7 Alarms menu Energy EM meters only...
Страница 66: ...IM P198 05 MI Issue 2 66 4 14 Diagnostics menu...
Страница 132: ...IM P198 05 MI Issue 2 132...