6.3 Level two hidden diagnostics values
4-20(1) Zero
Analogue counts to calibrate zero on Analogue output 1.
4-20(1) FScale
Analogue counts to cal. full scale on Analogue output 1.
4-20(2) Zero
Analogue counts to calibrate zero on Analogue output 2.
4-20(2) FScale
Analogue counts to cal. full scale on Analogue output 2.
4-20(3) Zero
Analogue counts to calibrate zero on Analogue output 3.
4-20(3) FScale
Analogue counts to cal. full scale on Analogue output 3.
Ext. 4 mA Cal.
Enter 0 for auto calibration or enter factory supplied A/D counts.
Note: You must connect a known 4.00mA input if you are going to
calibrate the unit.
Ext. 20 mA Cal.
Enter 0 for auto calibration or enter factory supplied A/D counts.
Note: You must connect a known 20.00 mA input if you are going
to calibrate the unit.
External Input
Enter what the external 4-20 mA input represents, i.e. Temperature
1, Temperature 2, or Pressure. The meter will use this for its internal
calculations.
Ext. Full Scale
Enter the full scale units that correlate to the 20 mA point. Note:
It must be in the units for the selected input type such as Deg F,
Deg C, Psi a, Bar A, etc.
Ext. Zero Scale
Same as above but for the 4 mA point.
Alarm (1) Test
Used as a test to verify that the alarm circuit is functioning. When
low is selected the alarm will initiate a low alarm on the output.
When High is selected it will give a high alarm on the output.
Alarm (2) Test
Used as a test to verify that the alarm circuit is functioning. When
low is selected the alarm will initiate a low alarm on the output.
When High is selected it will give a high alarm on the output.
Alarm (3) Test
Used as a test to verify that the alarm circuit is functioning. When
low is selected the alarm will initiate a low alarm on the output.
When High is selected it will give a high alarm on the output.
Reynolds Corr.
Reynolds number correction for the flow profile.
Gain Control
Manual gain control (factory use only). Leave set at 1.
Filter control
Manual filter control. This value can be changed to any number
to force the fi value to a constant. A value of zero activates the
automatic filter control which sets fi at a level that floats above
the f value.
High Pass Filter
Filter setting - Factory use only
Содержание RIM20
Страница 49: ...IM P198 05 MI Issue 2 49 Run mode screens Press Exit to return to Run mode Use keys to access each item...
Страница 50: ...IM P198 05 MI Issue 2 50 4 3 Using the set up menus Run mode screens...
Страница 52: ...IM P198 05 MI Issue 2 52 4 5 Output menu...
Страница 56: ...IM P198 05 MI Issue 2 56 4 7 Alarms menu Energy EM meters only...
Страница 66: ...IM P198 05 MI Issue 2 66 4 14 Diagnostics menu...
Страница 132: ...IM P198 05 MI Issue 2 132...