IM-P198-05
MI Issue 2
28
3.11 Display / Keypad adjustment (All meters)
The orientation of the display / keypad may be changed in 90 degree increments for easier
viewing.
Fig. 12 Display / keypad viewing adjustment
The electronics boards are electro-statically sensitive. Wear a grounding wrist strap and
make sure to observe proper handling precautions required for static-sensitive components.
To adjust the display:
1.
Disconnect power to the flowmeter.
2.
Loosen the small set screw which secures the electronics enclosure cover. Unscrew and
remove the cover.
3.
Loosen the 4 captive screws.
4.
Carefully pull the display/microprocessor board away from the meter standoffs. Make
sure not to damage the connected ribbon cable.
5.
Rotate the display/microprocessor board to the desired position. Maximum turn, two
positions left or two positions right (180°).
6.
Align the board with the captive screws. Check that the ribbon cable is folded neatly
behind the board with no twists or crimps.
7.
Tighten the screws. Replace the cover and set screw. Restore power to the meter.
Rotate display /
keypad in 90 degrees
increments (maximum
180 degrees from
original position)
Содержание RIM20
Страница 49: ...IM P198 05 MI Issue 2 49 Run mode screens Press Exit to return to Run mode Use keys to access each item...
Страница 50: ...IM P198 05 MI Issue 2 50 4 3 Using the set up menus Run mode screens...
Страница 52: ...IM P198 05 MI Issue 2 52 4 5 Output menu...
Страница 56: ...IM P198 05 MI Issue 2 56 4 7 Alarms menu Energy EM meters only...
Страница 66: ...IM P198 05 MI Issue 2 66 4 14 Diagnostics menu...
Страница 132: ...IM P198 05 MI Issue 2 132...