76
Calibration
CAL-V
TM
The T=xxx is a count down timer indicating how much time is left to finish
the test. A “Please Wait” message will be flashing on and off on line 2
during this test.
F1 F2 F3 F4
Upon test completion, the final CAL-V™ value will be displayed along with
a Pass/Fail message.
Performing the Zero CAL-CHECK™ Calibration Validation Test
The Zero CAL-CHECK™ test is a companion test to CAL-V™. Unlike
CAL-V™, which may be performed in the pipe and at process conditions,
Zero CAL-CHECK™ must be performed at zero flow to ensure a valid
test result. This test is used to confirm that the flow meter still retains
its original NIST-traceable calibration at zero flow and that the sensor is
free of film or residue tat may affect readings. The test takes less than 5
minutes to complete. At the conclusion of the test, a Pass or Fail message
will be displayed. Press
F4
at the conclusion of the test to return to normal
measuring mode or to terminate the test.
Press
FLO (F2)
from the main menu. The display will show:
F1 F2 F3 F4
Press
DGN (F1).
The display will show:
F1 F2 F3 F4
Zero CAL-
CHECK™
CAL-V = 0.51
Passed OK
FLOW PARAMETER 1
DGN UNT PRM EXIT
DIAGNOSTIC
SIM TST EXIT