Upon test completion, the final CAL-V™ value will be displayed along
with a Pass/Fail message.
Zero CAL-CHECK™ - Calibration Validation Test 2
The Zero CAL-CHECK™ test is a companion test to CAL-V™.
Unlike CAL-V™, which may be performed in the pipe and at process
conditions, Zero CAL-CHECK™ must be performed at zero flow to
ensure a valid test result. This test is used to confirm that the flowmeter
still retains its original NIST-traceable calibration at zero flow and that
the sensor is free of film or residue that may affect readings. The test
takes less than 5 minutes to complete. At the conclusion of the test, a
Pass or Fail message will be displayed. Press
F4
at the conclusion of
the test to return to normal measuring mode or to terminate the test.
See Calibration Validation User's Guide for more details.
Press
FLO (F2)
from the main menu. The display will show:
Press
DGN (F1).
The display will show:
Press
TST (F2).
The display will show:
Press
ZRO (F2)
to enter the Zero CAL-CHECK™ menu.
If performing the test in the pipe, a "no flow" condition must be met. If
performing out of the pipe, the meter must be removed and the sensor
FLOW PARAMETER 1
DGN UNT PRM EXIT
DIAGNOSTIC
SIM TST EXIT
CAL-V = 0.51
Passed OK
DIAGNOSTIC
CALV ZRO EXIT
52
Operation: Programming
Programming
Using
the Local Display