3124.6 SCFM
csv=0.3432 Volt
Enter: Press F1 & F2 at the same time
Press F4 to return to normal mode
Display 10
CsvAv=366809
Vel=112345.7 FT/M
Display 11
FloFlt=3666805.3
Vel=2356.45 M/H
Display 12
TsiAvr=512.5 cnt
TsvAvr=323.7 cnt
Display 13
Tsi= 2.1345 Volt
Tsv=0.9856 Volt
Display 14
Tsi = 0.0435 Amp
Tsi = 221.5 Ohm
Display 15
RTD9= 345.5 cnt
Gas Temp=123.7 °C
Display 16
CH1_420=2167 cnt
CH2_420=1234 cnt
Display 17
Feq=1234.5 cnt
Alarm=33,35
Display 18
FloHi= 1234 SCFM
FloLo=0 SCFM
Display 19
TmpHi=300 °C
TmpLo=10 °C
Display 20
Elp=12.5 HR
Stat(hex)=2800
Display 21
FT3 V3.02d
Display V2.03b
Display 22
Pwr_Cycl=24
Err_tot=0
Display 23
Tsi=221.5 Ohm
RTD9=10.3 Ohm
Display 24
CAL-V=23.51
CAL-V Chk=0.2%
Display 25
F2 Key
F1 Key
BrShtDnCnt=0 cnt
Display 26
F3 & F4 pressed at the same time will
initiate a "Total" reset
Flow in selected unit
Sensor voltage in volts
Sensor average volts
Velocity in selected unit
Sensor filtered average in volts
Velocity in meters/hour
TSI average count
TSV average count
TSI in volts
TSV in volts
TSI current in amps
TSI resistance in ohms
RTD9 count
Gas Temperature in degrees C
CH1 4-20ma current loop count
CH2 4-20 ma current loop count
Frequency output count
Alarm codes
High flow limit alarm
Low flow limit alarm
High temperature limit alarm
Low temperature limit alarm
Elapsed time in hours
Status in hexadecimal
MTI10/MTL10 main board firmware revision
MTI10/MTL10 display board firmware revision
Power cycle count
Error with totalizer count
TSI resistance in ohms
RTD9 resistance in ohms
CAL-V Value
CAL-V last verify value
Bridge shutdown detection count
ZRO_Pref=xx.xxxx
ZRO_diff=x.xx%
ZRO_Bref=xx.xxxx
ZRO_diff=x.xx%
Display 27
Display 28
Zero CAL-CHECK Pipe Ref
Zero CAL-CHECK % difference
Zero CAL-CHECK Bottle Ref
Zero CAL-CHECK % difference
MTI10/MTL10 Engineering Displays
Pressing the F1 & F2 keys at the same time in the normal mode, brings
up the engineering displays. These displays show internal parameters
of the MTI10/MTL10 which are used by Spirax Sarco service
technicians.
Press F4 to exit. Use the F1 & F2 keys to navigate.
Fig. 4.3: MTI10/MTL10 Engineering Displays
27
Operation: Engineering Displays
Engineering
Display