ST2516 Series Operation Manual
65
Time
Minimum value
Maximum value
t1: trigger pulse width
1ms
---
t2: measurement time at one
time
t3+t4
---
t3: sampling time of one
measurement
1 Sampling Time
---
t4: data processing and display
time of one measurement
Display “ON”: 22ms
Display “OFF”: 5ms
---
t5: end of sampling to control
output time
2ms
---
t6: measurement delay time
See the setup for
measurement delay
---
When OVC is OFF: Sampling Time = 5ms (50 Hz) / 5ms(60 Hz) Fast
20 ms (50 Hz) / 16.6ms(60 Hz) Medium
110 ms (50 Hz) / 110ms(60 Hz) Slow
450 ms (50 Hz) / 450ms(60 Hz) Slow
When OVC is ON: Sampling Time = 10 ms + t6 (50 Hz) / 10ms + t6 (60 Hz) Fast
40 ms + t6 (50 Hz) / 33ms + t6 (60 Hz) Medium
220 ms + 9* t6 (50 Hz) / 220 ms + 11* t6 (60 Hz) Slow
900 ms + 39*t6(50 Hz) / 900 ms + 47*t6(60 Hz) Slow