ST2516 Series Operation Manual
32
5.1.6
Resistance Measurement Time
When OVC is OFF, measurement time= 5ms +t1 (50 Hz) / 5ms+t1 (60 Hz) FAST
20ms+t1 (50 Hz) / 16.6ms+t1 (60 Hz) MED
110ms+t1 (50 Hz) / 110ms+t1 (60 Hz) SLOW
450ms +t1 (50 Hz) / 450ms+t1 (60 Hz) SLOW
When OVC is OFF, measurement time= 10 ms + t+t1 (50 Hz) / 10ms + t+t1 (60 Hz) FAST
40ms + t+t1 (50 Hz) / 33ms + t+t1 (60 Hz) MED
220ms + 9* t+t1 (50 Hz) / 220 ms + 11* t+t1 (60 Hz) SLOW
900ms + 39*t+t1 (50 Hz) /900 ms + 47*t+t1 (60 Hz) SLOW
NOTE: t is the measurement waiting time, t1 is the data processing time, approximately is 5ms.
More detailed information about time parameters, please refer to Chapter 7 “Handler Interface”.
5.1.7
Average
Range from 1 to 255, programmable: this value reflects the measurement times from measuring
resistance to measuring display.
5.2
Test Signal
5.2.1
Current range
ST2516: Current range:
1μA
-- 1A
ST2516A: Current range: 10
μA
– 100mA
ST2516B: Current range: 100
μA
-- 1A
5.2.2
Output Voltage of Open Circuit
Output voltage of open circuit: 0.7V, 3V, 40mV
5.2.3
Maximum Display Range
Parameter
Measurement Display Range
R
1u
Ω
–- 2
MΩ
LPR
0.1m
Ω
—2.1k
Ω
T
-99.9
℃
-- 999.9
℃