ST2516 Series Operation Manual
18
status as OFF.
3.5
<MEAS SETUP>
Touch the main menu. And then touch the soft key MEAS SETUP to enter into the MEAS SETUP
page shown as figure 3-5.
FILE
MORE 2/2
11:36:54
<MEAS SETUP>
FUNC
R-T
RANGE
AUTO
TRIG
INT
TRIG DELAY
AUTO
AVERAGE
56
SPEED
MED
0 ADJ
ON
OVC
OFF
MEAS
SETUP
TC/
Δ
t
SETUP
BIN
SETUP
SYSTEM
SETUP
MED
Figure 3-5 Measurement Setup
On this page, the following parameter can set:
FUNC R-T
: Touch this key to select the required parameter.
RANGE AUTO
: Touch this key, the following soft keys will be displayed: AUTO (automatically
set the resistance range), HOLD (lock the resistance range), ↑ (increase the range), ↓
(decrease the range). Touch the corresponding soft key to set the range mode.
TRIG INT
: Touch this key, the following soft keys will be displayed in the soft key zone: INT
(auto measurement mode), MAN (manual measurement mode), EXT (external trigger mode),
BUS (BUS trigger mode).
TRIG DELAY
: Touch this key, the following soft keys will be displayed:
AUTO
,
MANUAL
,
INPUT
.
NOTE
: The setting of manual measurement delay time is determined by inputting. Range:
0ms—9.99ms. If the delay time is set as 0ms, the detection for wrong measurement cannot
be executed. Therefore, it is recommended to set the delay time being more than 1ms.
AVERAGE
: Touch this key, a numeric keyboard will pop up. Input the average number
ranging from 1 to 255. The larger the average number is set, the better the accuracy will be
obtained but the longer time it will cost to display the result.
SPEED MED
: This mode is usually applied in 4-terminal measurement. Touch this key,
SLOW1, SLOW2, FAST
and
MED
will be selectable. In
SLOW1
and
SLOW2
mode, the noise is
lower; in FAST mode, noise suppression is needed.
FAST
is high speed measurement.
0 ADJ:
Touch this key to select
ON
or
OFF
.
ON
is to turn on the short correction function
while
OFF
is to turn off.
OVC:
Touch this key to select
ON
or
OFF
.
ON
is to turn on the compensation voltage while
OFF
is to turn off.