ST2516 Series Operation Manual
40
pattern”, “self-calibration mode”, etc.. The DISPlay:PAGE? qurey returns to the current page.
Command Tree:
The :IMPedance command sets the “function” parameters.
The FUNCtion:IMPedance? query returns to the current “function” parameters.
Command Syntax: FUNCtion:IMPedance <function>
Details are as follows:
R Set “function” as R.
RT Set “function” as R-T. (ST2516)
T Set “function” as T. (ST2516)
LPR Set “function” as LPR.
LPRT
Set “function” as LPR-T. (ST2516)
For example: WrtCmd(“FUNC:IMP RT”); Set “function” as R-T .
Query syntax: FUNCtion:IMPedance?
Return Format: <function><NL^END>
If the current “function” is R-T, it will returns to RT.
The :IMPedance:RES:RANGe command sets the instrument range in common resistance
measurement pattern.
The FUNCtion:IMPedance:RES:RANGe? query returns to the current range parameters in
common resistance measurement pattern.
Command Syntax: FUNCtion:IMPedance:RANGe <value>
Where, <value> can be the impedance of the tested LCR component or the specific
range values. Its data format is NR1, NR2, NR3, <value> = 0 to 2E+6.
For example: WrtCmd(“FUNC:IMP:RES:RANG 123”); Set the instrument range in common