ST2516 Series Operation Manual
10
R (Resistance)
R-T (Resistance and temperature)
T (Temperature)
LPR (resistance test at low current mode)
LPR-T (Temperature and resistance test at low current mode)
----------------------------------------------------------------------------------------------------------------------
Note: Settings and modifications of all functions can be taken by the following steps:
1.
Touch the corresponding function zone and then select the desired soft keys.
2.
Use arrow buttons and [ENTER] button to set and modify the corresponding function.
----------------------------------------------------------------------------------------------------------------------
Setting steps for measurement function:
You use buttons or touch the screen to select
FUNC
, available soft keys will be displayed in the
right soft key zone.
R
R-T (ST2516)
T (ST2516)
LPR
LPR-T (ST2516)
3.1.2
Measurement Range
There are two resistance modes: resistance measurement mode and resistance measurement at
low voltage mode. Measure and display two types of parameters: resistance parameters and
temperature parameters.
ST2516
has 9 DC resistance ranges:
20mΩ, 200mΩ, 2Ω, 20Ω,
200
Ω
,
2kΩ, 20kΩ,
2
00kΩ,
2
MΩ
ST2516A
has 7 DC resistance ranges:
200mΩ
,
2Ω
,
20Ω
,
200Ω
,
2kΩ
,
20kΩ
, 200k
Ω
ST2516B
has 7 DC resistance ranges:
20mΩ
,
200mΩ
,
2Ω
,
20Ω
,
200Ω
,
2kΩ
,
20kΩ
ST2516, ST2516A, ST2516B
has 4 DC low voltage resistance ranges:
2Ω, 20Ω, 200Ω, 2KΩ
ST2516
The testing range of temperature (PT500):
-10
℃ to 99.9℃.
The testing range of temperature (Analog):
0 to 2V
Operation steps for setting measurement ranges:
1)
Touch the range zone, the following soft keys will be displayed.
AUTO
Set the range mode as AUTO.
HOLD
Switch the range mode from AUTO to HOLD. When the range mode is set as HOLD,
the range will be locked at the current measurement range which is displayed in the
RANG
zone.
↑
(+)
Increase the range.
↓
(-)
Decrease the range.
2)
Touch the corresponding soft key to select the required range.