
- 44 -
SORHEA
G-WALL
Setting the sensitivity:
The value of sensitivity determines the threshold of alarm triggering.
When an alarm is generated, the shock value is recorded in the log history.
The table below gives the possible values that can be found in the log history function and the value
of the selected sensitivity.
Shock values displayed in the log history
16
15
14
13
12
11
10 9 8 7 6 5 4 3 2 1
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
value of sensitivity
1
Example: For a shock of a value of 7:
If the sensitivity is set to 16, the display will be in the log history "saturation"
If the sensitivity is set to 10, the display will be in the log history "7"
If the sensitivity is set to 5, no alarm will be triggered.
7.3 Exit the settings mode
Move the switch to “DETECT”.
The buzzer will emit 2 short beeps to confirm exit into detection mode.
The shock is less than the value of the selected sensitivity, no alarm trigger
The shock triggers an alarm.
The shock triggers an alarm but the shock value is in saturation.
Saturation zone
No alarm
triggered