CSM-200
7-3
3.
Enter the MMT display in the administrator mode.
4.
Select “Program” of “Test” menu and open the Program display.
5.
Input “test” into the Command box and set MMT into test mode.
→
When test mode is established, the MMT cannot be controlled any more from the host system.
To exit the test mode, input “remote” into the Command box, or close the Program display.
6.
Input “init” into the Command box.
→
System initialization of the PetaSite starts.
→
If any trouble occurs during initialization, check error code of the PetaSite, perform the recovery
operation and initialize the PetaSite again.
The error code contents and the suspected cause of error can be known using the help menu of
error code.