Transducer Checks
The transducer checks option provides the ability to run the DVM, scope or spectrum
displays or perform calibration checks. Select the desired channel or channels from
the list and the preferred option using the radio buttons and then click Run to execute
the option for the selected transducer or transducers.
Rezero Display
Use the rezero display to view and rezero the defined transducer channels. Select the
desired channel or channels and click OK. TCE continuously reads and displays the
current transducer signals in engineering units. Select the hold option to pause display
updates or use the scroll bar at the right of the window to adjust the display update
rate. The rezero display also shows the prerun rezero value defined for the transducer
in the test setup and indicates in the (A) column if the transducer is to be rezeroed just
prior to the next test run. To manually rezero a channel, select the transducer in the
(B) column and click the rezero button or press ALT+R.
4.3.3
Previewing a Test Run
A preview run effectively starts a test run with all DataMode storage suppressed. All of
the transducer channels and computed channels run in preview mode just as they
would during a data collection run. To start a preview run, select Preview Run from the
Test Control menu or toolbar.
4.3.4
Starting a Test Run
To start a test run, select Start Run from the Test Control menu or toolbar. The Somat
eDAQ
lite
supports multiple test runs for an initialized test. Start a new test run anytime
after a previous run stops or during a preview run. When starting a test run, TCE
presents a start test run dialog window displaying the next run number and allowing
the input of a short run description (up to 63 characters per run).
NOTE
Use TCE General Preferences to suppress the start test run dialog window.
Option
Description
Calibration Check
Run a calibration check. Note that the calibrations are the current
transducer calibrations which may differ from the calibrations defined
at test initialization if the transducer has been rezeroed. Check for this
difference using the rezero offset option.
DVM Display
Run the DVM display.
Scope Display
Run the scope display.
Spectrum Display
Run the spectrum display.
Rezero Offset
View the differences between the current transducer calibration
intercepts and those defined at test initialization. For each transducer
channel, this difference is the cumulative sum of all rezero offsets in
engineering units since test initialization.
HBM: public
Содержание eDAQlite
Страница 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Страница 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Страница 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Страница 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Страница 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Страница 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Страница 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Страница 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Страница 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Страница 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Страница 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...
Страница 227: ...I2773 4 4 en 227 Somat eDAQlite HBM public ...