NOTE
Always provide the +12 volt REF voltage for the SWC interface. For other CAN
interfaces, the red power wire can be used to power transducers with a
user-configurable range of 3 to 24 volts.
12.3
ELHLS (High Level Analog Layer)
ELHLS Transducer Cable
The ELHLS layer uses a Somat SAC-TRAN-MP Transducer Cable
(1-SAC-TRAN-MP-2-2 or 1-SAC-TRAN-MP-10-2) with an M8 connector and a set of
color-coded pigtail wires. The following table lists the pinouts for the SAC-TRAN-MP
cable when used for an ELHLS input.
12.4
ELBRG (Bridge Layer)
ELBRG Transducer Cable
The ELBRG layer uses a Somat SAC-TRAN-MP Transducer Cable (SAC-TRAN-MP)
with an M8 connector and a set of color-coded pigtail wires. The following table lists
the pinouts for the SAC-TRAN-MP cable when used for an ELBRG input.
Function
Pin
Wire Color
SWC
1
Brown
CANH
2
White
AGnd
3
bare wire
Power
5
Red
+12 V REF
5
Red (SWC)
CANL
6
Green
Function
Pin
Wire Color
reserved
1
Brown
+ Signal Input
2
White
Shield
3
bare wire
Ground
4
Black
Power
5
Red
- Signal Input
6
Green
Function
Pin
Wire Color
reserved
1
Brown
+ Signal Input
2
White
Shield/Ground
3
bare wire
HBM: public
Содержание eDAQlite
Страница 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Страница 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Страница 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Страница 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Страница 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Страница 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Страница 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Страница 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Страница 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Страница 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Страница 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...
Страница 227: ...I2773 4 4 en 227 Somat eDAQlite HBM public ...